Service

 
 

Sophisticated Analytical Instrumentation Facility, CIL and UCIM

Atomic Absorption Spectroscopy

 

SPECIFICATIONS:
Monochromator: Mounting Littrow type, Focal Length: 250 mm, Diffraction Grating: 1800 Lines/mm, Plane, 30 x 30 mm, Grating Blazed (at 250 nm Bandwidth: Continuously Variable from 0.1 to 2.0 nm with 0.1 nm Steps)

Flame atomizer

  • Auto-ignition of flame following all safety interlock checks:
    • Flame door lock
    • Burner presence
    • Burner type sensor
    • Drain flow sensor
    • Lamp door
    • Pressure sensor
  • Inert nebulizer and spray chamber suitable for all types of samples prepared with various acids

Key Hardware

  • Double beam optics to improve signal stability
  • Automatic fuel flow and burner position optimisation to ease method development
  • Flexibility and reduced warm-up time for lamps, increasing productivity
  • Auto detection of coded lamps
  • Magnetic lock enabled doors with safety sensors

SAMPLE REQUIREMENTS:
Lamps Available for Analysis: Ca, Cu, Zn, Mn, K, Br, Mg, Fe
10 ml clean solution water base sample

Contact Person: Raj Kumar Sharma 9876132758  

 

Enquiry No(s): 1800-180-2065, +91 172 2534818, 2534866 (from 9:30 am to 1:00 pm & 2:30 pm to 5:00 pm on working days )